Sign In | Join Free | My burrillandco.com
Home > Patient Monitor Accessories >

10097559 PSR-11-75-KE13PB Oxygen Sensor For Covidien PB760 PB840 PB980 Puritan Bennett Ventilator Machine

China World Technology Medical Equipment Service Group
Trust Seal
Verified Supplier
Credit Check
Supplier Assessment
Contact Now
    Buy cheap 10097559 PSR-11-75-KE13PB Oxygen Sensor For Covidien PB760 PB840 PB980 Puritan Bennett Ventilator Machine from wholesalers
     
    Buy cheap 10097559 PSR-11-75-KE13PB Oxygen Sensor For Covidien PB760 PB840 PB980 Puritan Bennett Ventilator Machine from wholesalers
    • Buy cheap 10097559 PSR-11-75-KE13PB Oxygen Sensor For Covidien PB760 PB840 PB980 Puritan Bennett Ventilator Machine from wholesalers
    • Buy cheap 10097559 PSR-11-75-KE13PB Oxygen Sensor For Covidien PB760 PB840 PB980 Puritan Bennett Ventilator Machine from wholesalers
    • Buy cheap 10097559 PSR-11-75-KE13PB Oxygen Sensor For Covidien PB760 PB840 PB980 Puritan Bennett Ventilator Machine from wholesalers
    • Buy cheap 10097559 PSR-11-75-KE13PB Oxygen Sensor For Covidien PB760 PB840 PB980 Puritan Bennett Ventilator Machine from wholesalers

    10097559 PSR-11-75-KE13PB Oxygen Sensor For Covidien PB760 PB840 PB980 Puritan Bennett Ventilator Machine

    Ask Lasest Price
    Brand Name : Covidien
    Model Number : 10097559 PSR-11-75-KE13PB
    Payment Terms : T/T, Western Union,Paypal
    Delivery Time : 3-5 working days
    MOQ : One piece
    • Product Details
    • Company Profile

    10097559 PSR-11-75-KE13PB Oxygen Sensor For Covidien PB760 PB840 PB980 Puritan Bennett Ventilator Machine

    10097559 PSR-11-75-KE13PB oxygen sensor for Covidien PB760 PB840 PB980 Puritan Bennett Ventilator Machine


    Oxygen sensor calibration test


    Test the oxygen sensor calibration as follows:


    1. Connect the ventilator oxygen tubing to a known source of 100% O2 (e.g., a medical oxygen cylinder). Press the 100% O2/CAL button for 2


    minutes or the INCREASE O2 button for 2 minutes to calibrate the oxygen sensor. Once the button light turns off, proceed to the next step.


    2. Connect the ventilator oxygen tubing to another known source of 100% O2 (e.g., a second medical oxygen cylinder).


    3. Set %O2 to the following values ​​and wait one minute for each monitored value to stabilize: 21% 40% 90%


    4. Observe the upper display to ensure that the O2 value (% O2 delivered) is within 3% of each setting within one minute of selecting each


    setting.


    Oxygen Sensor Calibration Button: Older ventilators use the 100% O2/CAL 2 minutes button, while newer ventilators use the INCREASE O2 2


    minutes button. Provides 100% oxygen (if available) for two minutes and calibrates the oxygen sensor. The green light on this button will


    illuminate and a message (100% O2 Cal in Progress) will appear on the lower touchscreen to indicate 100% O2 delivery is active. If you press the


    O2 button again, the system will restart the two-minute delivery interval. Press CANCEL to stop calibration. For information on calibrating the


    oxygen sensor, see page TR 15-5.


    Alarm Messages


    If you see this O2 Sensor message, it means that background checks have identified a problem with the oxygen sensor (sensor error or not calibrated). Patient ventilation will not be affected. And you can:


    • Press 100% O2 CAL or O2 INCREASE for 2 minutes to recalibrate the oxygen sensor.


    • Disable the oxygen sensor.


    • Replace the oxygen sensor.


    Specification of 10097559 PSR-11-75-KE13PB oxygen sensor for Covidien PB760 PB840 PB980 Puritan Bennett Ventilator Machine


    Product Name

    PSR-11-75-KE13PB 10097559 oxygen sensor for Covidien PB760 PB840 PB980 Puritan Bennett Ventilator Machine


    BrandCovidien
    Part No

    PSR-11-75-KE13PB 10097559


    Application

    PB760 PB840 PB980 Puritan Bennett Ventilator Machine


    MOQOne piece
    ConditionOriginal /Compatible new
    Warranty90 days

    Photo


    Below photo is PSR-11-75-KE13PB 10097559 oxygen sensor ,which can be used for Covidien PB760 PB840 PB980 Puritan Bennett Ventilator Machine .For more details ,please feel free to contact us .



    Thank you so much for your attention .If you're interested in 10097559 PSR-11-75-KE13PB Oxygen Sensor For Covidien PB760 PB840 PB980 Puritan Bennett Ventilator Machine ,feel free to contact us .Thank you .

    Quality 10097559 PSR-11-75-KE13PB Oxygen Sensor For Covidien PB760 PB840 PB980 Puritan Bennett Ventilator Machine for sale
    Inquiry Cart 0
    Send your message to this supplier
     
    *From:
    *To: China World Technology Medical Equipment Service Group
    *Subject:
    *Message:
    Characters Remaining: (0/3000)