Zoll E Series Defibrillator Main Board 9301-1380 9300-1380 REV A
► Main System Board Functions
The main motherboard contains the main computing and control elements of the E Series device. The Printed Wiring Board (PWBA) receives
signals from the front panel control switches, ECG input connectors, and functional modules such as the AC/DC battery charger.
pacemaker/defibrillator module, the strip chart recorder and the PCMCIA card interface and (if applicable) such). ) SpO2, EtCO2 and NIBP
modules. The main motherboard monitors and processes these input signals to generate other signals that: 1) control the operation of other
modules in the system; 2) controls the front panel display and audio output of the device; and 3) store data for retrieval using summary reports,
PCMCIA cards, and/or modem.
The main board electronics and software provide the following primary functions of the E Series:
1. Main processor and memory.
2. Acquisition of ECG signal and processing of ECG leads, including:
• A/D conversion.
• Detection of ECG signals.
• QRS detection.
• Detection of an implanted pacemaker.
• Heart rate counting.
• Shockable ECG rhythm analysis.
3. Data communication with SpO2, EtCO2 and NIBP modules and their control.
4. Safety control and monitoring of pacemaker and defibrillator functions provided by high voltage module and biphasic bridge module.
5. Treatment of physiological alarms.
6. Monitor the front panel control switch, monitor the connection of accessories, and monitor the system response to activating the switch or connecting E-Series accessories.
7. Format and refresh the front panel display.
8. The main power supply of the device.
9. Generation and control of sound output (e.g. alarms, voice commands, warning sounds).
10. Sound signal processing, data compression, sound recording data storage.
11. Real-time clock and other timing functions.
12. Summary report, 12-lead ECG report data storage.
13. Monitor battery status and control battery charging functions performed by PWBA AC or DC charger.
14. Data transfer and control includes system interconnection PCB functions, including tape diagram recorder and PCMCIA slot functions.
► Specification of Zoll E Series Defibrillator Main Board 9301-1380 9300-1380 REV A
Product Name | Zoll E Series Defibrillator Main Board 9301-1380 9300-1380 REV A |
Brand | ZOLL |
Part No | 9301-1380 9300-1380 REV A |
Compatible with | Zoll E Series Defibrillator |
MOQ | One piece |
Warranty | 90 days |
Thank you so much for your attention .If you're interested in Main Board 9301-1380 9300-1380 REV A For Zoll E Series Defibrillator ,feel free to contact us .Thank you .
![]() |